====== Requirements ====== These are suggestions, for people who want numbers... ===== Room ===== The microscope should be installed in a room with a closable door that is climate controlled. Room lights should be turned off during microscope operation. There should be 0.5 m of space on all sides of the air table. The acquisition computer should be set up on a separate desk next to the air table. It is recommended to have shelves either above the air table behind the user-facing direction or else underneath the air table for placing the laser, controller, etc. ===== Air Conditioning ===== The room temperature should be controlled within +/- 0.5 °C at all times. If the room deviates from this, 5 hours of stabilization is recommend for every °C of deviation. Having an enclosure around the microscope can mitigate effects of changing room temperature. AC cooling capacity of 1000 W is usually adequate (depends on cameras and laser). The microscope should be placed away from vents or other sources of air flow. ===== Acoustic Noise ===== The microscope should be installed in a quiet room; sounds waves are vibratory and can couple to the instrument and impact image quality. It is recommended to have noise < 50 dB where the instrument is placed. ===== Vibration ===== The microscope should be installed on an air table with at least 90% isolation at 10 Hz (-20 dB). The smallest recommended air table is 750 x 900 mm. ===== Electromagnetic ===== The microscope does not normally require any electromagnetic shielding. ===== AC Power ===== About 10 standard AC plugs are required depending on the specifics of the microscope. It is recommended to have dedicated outlets for the laser launch and computer; other items can be on power strips.