Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision
Previous revision
hardware:controller [2017/08/02 16:01]
Jon Daniels [Required software and instructions]
hardware:controller [2024/02/16 17:54] (current)
Jon Daniels [Firmware]
Line 8: Line 8:
 In general firmware updates add new functionality and fix bugs while retaining backwards compatibility, though on rare occasion backward compatibility is compromised.  If you are not using Micro-Manager, you should double-check with the vendor of the acquisition software before updating firmware.  If you are using Micro-Manager, you should generally download the latest nightly build whenever you update firmware.  Conversely, if you are having problems with a recent build of Micro-Manager make sure that the controller firmware is the latest. In general firmware updates add new functionality and fix bugs while retaining backwards compatibility, though on rare occasion backward compatibility is compromised.  If you are not using Micro-Manager, you should double-check with the vendor of the acquisition software before updating firmware.  If you are using Micro-Manager, you should generally download the latest nightly build whenever you update firmware.  Conversely, if you are having problems with a recent build of Micro-Manager make sure that the controller firmware is the latest.
  
 +** Warning: these instructions are for iSPIM and diSPIM only.  If you have another type of ASI light sheet microscope (e.g. oSPIM or single-objective light sheet) contact ASI for guidance and for the proper firmware files.  **
 ==== Before you begin ==== ==== Before you begin ====
  
Line 15: Line 16:
 ==== Required software and instructions ==== ==== Required software and instructions ====
  
-To update the firmware you need the [[https://www.dropbox.com/s/vt2mq97kaenzpl4/ASITigerConsole-1.1.0.zip?dl=1|ASI Tiger Console]] software and the [[https://www.dropbox.com/sh/r5cabcfytu0e88f/AAD3IlCEpKInWmm2h4oh5_z6a?dl=1|latest diSPIM firmware package from ASI]] (last updated 2017-02-08, firmware version 3.17), or from the [[https://www.dropbox.com/sh/tpl82yahjz41nil/AABnSsCRDR_gRhxTiOi0srUJa?dl=0|diSPIM firmware archive]].  Use [[http://7zip.org|7zip]] to decompress. This firmware should work for diSPIM systems sold 2014 or later.  A link to download firmware for the oSPIM is found on the [[docs:ospim|oSPIM page]].+To update the firmware you need the [[http://www.asiimaging.com/index.php/support/downloads/tiger-controller-console/|ASI Tiger Console]] software and the [[https://www.dropbox.com/sh/r5cabcfytu0e88f/AAD3IlCEpKInWmm2h4oh5_z6a?dl=1|latest diSPIM firmware package from ASI]] (dated 2020-10-05, firmware version 3.33), or from the [[https://www.dropbox.com/sh/tpl82yahjz41nil/AABnSsCRDR_gRhxTiOi0srUJa?dl=0|diSPIM firmware archive]].  Use [[http://7zip.org|7zip]] to decompress. This firmware should work for diSPIM systems sold 2014 or later.  A link to download firmware for the oSPIM is found on the [[docs:ospim|oSPIM page]].
  
 General instructions for using the ASI Tiger Console software for updating the firmware are [[https://www.dropbox.com/s/j85783gltskreth/Using%20Dwayne%27s%20Tiger%20Console.pdf?dl=0|here]]. General instructions for using the ASI Tiger Console software for updating the firmware are [[https://www.dropbox.com/s/j85783gltskreth/Using%20Dwayne%27s%20Tiger%20Console.pdf?dl=0|here]].
  
-Standard firmwares for card addresses on diSPIM controllers are as follows.  Omit any cards that your controller doesn't have.  If your controller is otherwise non-standard (e.g. the addresses don't match the labels on the front of the controller) feel free to contact ASI for help.+Standard firmwares for card addresses on diSPIM controllers are as follows.  Omit any cards that your controller doesn't have.  Almost always you will be programming the same firmware build name as exists on your current controller and you can see the list if you are connected via serial to the controller when it powers up or else by issuing the [[http://asiimaging.com/docs/commands/who|WHO]] command.  If your controller is otherwise non-standard (e.g. the addresses don't match the labels on the front of the controller) feel free to contact ASI for help.
  
-^ Address  ^ Card          ^ Usual Firmware                   ^ Comments                                   +^ Address  ^ Card          ^ Usual Firmware                                ^ Comments                                                                                                             
-| 0/30     | Comm card     | TigerComm.hex                                                               +| 0/30     | Comm card     | TigerComm                                                                                                                                                          
-| 1/31     | XY stage      | SCAN_XY_LED%%__%%Adr1.hex        |                                            +| 1/31     | XY stage      | SCAN_XY_LED_SO%%__%%Adr1 (16 TPI stages)  | SCAN_XY_LED%%__%%Adr1 for 4TPI stages, "SO" version for scan-optimized stages which have been standard from mid-2016 on  
-| 2/32     | Z/F stages    | STD_ZF_LED%%__%%Adr2.hex         | STD_ZF%%__%%Adr2.hex if no LED being used  +| 2/32     | Z/F stages    | STD_ZF_LED%%__%%Adr2                      | STD_ZF%%__%%Adr2 if no LED being used, FTP_ZF if you have an FTP system (XYZ stage) instead of inverted microscope       
-| 3/33     | Micromirror   | MMIRROR_SPIM%%__%%ABCD_Adr3.H01                                             +| 3/33     | Micromirror   | MMIRROR_SPIM%%__%%ABCD_Adr3                                                                                                                                        
-| 4/34     | Piezo P       | ADEPT_SPIM%%__%%P_Adr4.H01                                                  +| 4/34     | Piezo P       | ADEPT_SPIM%%__%%P_Adr4                                                                                                                                             
-| 5/35     | Piezo Q       | ADEPT_SPIM%%__%%Q_Adr5.H01                                                  +| 5/35     | Piezo Q       | ADEPT_SPIM%%__%%Q_Adr5                                                                                                                                             
-| 6/36     | Progr. Logic  | PLOGIC_16%%__%%E_Adr6.hex                                                   |+| 6/36     | Progr. Logic  | PLOGIC_16%%__%%E_Adr6                     |                                                                                                                          | 
 +| 7/37     | MotIris     IRIS%%__%%TU_Adr7                           
  
 In general you should update all the firmware on your controller together.  One exception is filter wheel firmware which generally does not need updating. In general you should update all the firmware on your controller together.  One exception is filter wheel firmware which generally does not need updating.
Line 35: Line 37:
  
  
-==== Extra step with scan-optimized XY stages ====+==== Extra step with pre-3.20 XY stage firmware ====
  
-In late 2016 ASI transitioned to using scan-optimized XY stages with SPIM systems by default.  These stages have 16 TPI leadscrews instead of the usual 4 TPI leadscrew.  The leadscrew pitch setting is not stored across firmware updatesso after updating firmware if you have to tell the stage the correct leadscrew pitch (if it is not 4 TPI).  If you forget to do this, stages with 16 TPI leadscrew will move only quarter of the distance than the firmware thinks it is moving because there is essentially an extra four-fold gearing that the firmware doesn't know about.+In late 2016 ASI transitioned to using scan-optimized XY stages with SPIM systems by default.  These stages have 16 TPI leadscrews instead of the usual 4 TPI leadscrew, along with removing anti-backlash gearing in the X axis for smoother scanning. 
 + 
 +As of firmware version 3.20 (June 2018) there are two firmware versions available for the XY card and you should use the one that matches your stage.  Use SCAN_XY_LED_SO%%__%%Adr1.hex for scan-optimized 16 TPI stages (note the "SO") or SCAN_XY_LED%%__%%Adr1.hex  for 4 TPI stages. 
 + 
 +Using firmware prior to v3.20, after flashing the XY card firmware you need to set the leadscrew pitch using a serial command if you have a scan-optimized stage.  That way the controller knows that the leadscrew pitch is 16 TPI and can report distances accordingly.  Unless told otherwise the firmware assumes 4 TPI leadscrew
 + 
 +To tell the firmware ​about the leadscrew pitch you send single ​serial command ​to the controller.  ​In Micro-Manager you can use the property "​SerialCommand"of the TigerCommHub device or you can use some other program, both methods ​are described in [[http://​asiimaging.com/​docs/​tech_note_rs232_comm | ASI's serial communication documentation]]. ​  ​Send ​the serial command, then turn off the controller before doing anything else.  ​After powering the controller back on verify that the setting registered If you are using Micro-manager you will need to shut it down and restart as well.  For 16 TPI leadscrews use the following command: ​(others can be found in the [[http://​asiimaging.com/​docs/​commands/​customa|ASI documentation]]) 
 + 
 +''​**%%1 CCA X=6%%**''
  
-To tell the firmware about the leadscrew pitch you need to send a single serial command to the controller.  In Micro-Manager you can use the property "SerialCommand" of the TigerCommHub device or you can use some other program, both methods are described in [[http://asiimaging.com/docs/tech_note_rs232_comm | ASI's serial communication documentation]].   Send the serial command, then turn off the controller.  After powering the controller back on verify that the setting registered.  If you are using Micro-manager you will need to shut it down and restart as well.  For 16 TPI leadscrews use the following command: (others can be found in the [[http://asiimaging.com/docs/commands/customa|ASI documentation]]) 
  
-''**%%1 CCA X=6%%**'' 
  
  
Line 67: Line 75:
 ===== Stage Scanning ===== ===== Stage Scanning =====
  
-Instead of moving the sheet position and imaging piezo to create a stack of images, the sample can be translated through a fixed sheet.  Generally speaking this is best for large and/or flat samples.  The main advantages are that the light sheet can be thinner for flat samples because the beam waist stays at the same height relative to the coverslip and that there is a nearly unlimited stack "Z" (instead of limitation of piezo range).  The main disadvantages are that the output data need to be sheared or shifted before conventional analysis and the minimum time between successive time points is somewhat longer (because the motorized stage can't return to its starting position as fast as the piezo objective mover).  In the Micro-Manager implementation, the stage moves continuously in the X axis at an appropriate rate to give the user-specified slice spacing and the beam sweep creating the sheet creates a "stop motion" effect.  The stage speed is very uniform and the initial trigger is synchronized with the stage's encoder to give extremely repeatable positioning.+Instead of moving the sheet position and imaging piezo to create a stack of images, the sample can be translated through a fixed sheet.  Generally speaking this is best for large and/or flat samples.
  
-Stage scanning acquisition requires: +The main advantages of stage scanning are that the light sheet can be thinner for flat samples because the beam waist stays at the same height relative to the coverslip and that the "Z-stack" is limited only by the travel of the X-axis of the stage (instead of the piezo range).  The main disadvantages are that the output data need to be deskewed or sheared before conventional analysis and the minimum time between time points is somewhat longer than with a piezo. (Typically the stage is plenty fast for acquisition, i.e. the acquisition is camera/photon limited, but the motorized stage can't return to its starting position as fast as a piezo stage).  Users of static light sheets should also address PSF-blurring resulting from stage scanning; Nyquist demands >2 images per stationary PSF, but motion blur requires you to sample more.((The concept of blurred PSFs due to stage scanning is discussed in Chapter 3, Section 5 of [[http://doi.org/10.25560/68022|Vincent Maioli's PhD thesis]] from the Dunsby lab.))  The swept beam of a digital light sheet produces a "stop motion" effect and therefore does not require increased sampling. 
-  - An ASI-made XY stage, whether on a RAMM or other inverted microscope (requires tight synchronization of the stage with the light sheet/laser/camera).  Since late 2016 ASI has generally shipped scan-optimized stages with SPIM systems; these stages have 16 TPI leadscrews for a slower overall motion (and hence better accuracy with slow moves) and omit the anti-backlash gearing on the X axis for smoothest possible scanning.+ 
 +In the usual implementation, the stage moves continuously in the X axis at an appropriate rate to give the user-specified slice spacing.  The speed of the scan optimized stages is very uniform (see hardware notes below) and the stage's encoder triggers the acquisition internally to give extremely repeatable positioning of repeated stacks. 
 + 
 + 
 +==== Requirements for Stage Scanning ==== 
 + 
 +If you have a SPIM system 2016 or later you should be good to go.  If you have an earlier system you should verify the following if you have any trouble with stage scanning: 
 + 
 +  - An ASI-made XY stage, whether on a RAMM or other inverted microscope (requires tight synchronization of the stage with the light sheet/laser/camera).  In mid 2016 ASI transitioned to "scan-optimizedstages with SPIM systems; these stages have 16 TPI leadscrews for a slower overall motion (and hence better accuracy with slow moves) and omit the anti-backlash gearing on the X axis for smoothest possible scanning.
   - "SCAN" firmware module on XY card in the Tiger controller   - "SCAN" firmware module on XY card in the Tiger controller
-    * Most users should use SCAN_XY_LED firmware on the XY card (this is the same as STD_XY_LED except with scan module present); this has been in the firmware releases for download (on this page [[hardware:controller#firmware|above]]) since late 2015 +    * Use SCAN_XY_LED firmware variant on the XY card; this has been in the firmware releases for download (on this page [[hardware:controller#firmware|above]]) since late 2015
-    * All firmware should be updated to v3.10 or later (3.09 may be OK)+    * As of June 2018 there are two firmware versions available and flash the one that matches your stage. Use SCAN_XY_LED_SO%%__%%Adr1.hex for scan-optimized 16 TPI stages (note the "SO") or SCAN_XY_LED%%__%%Adr1.hex for 4 TPI stages. 
 +    * All firmware should be updated to v3.10 or later
   - XY card Rev F2 or later with proper jumper (standard after Jun 2015)   - XY card Rev F2 or later with proper jumper (standard after Jun 2015)
     * Make sure the following jumper is present: {{:hardware:stage_scan_xy_jumper.png?linkonly|this photo}}     * Make sure the following jumper is present: {{:hardware:stage_scan_xy_jumper.png?linkonly|this photo}}
Line 81: Line 98:
   - Software support (present in Micro-Manager plugin since mid-2015)   - Software support (present in Micro-Manager plugin since mid-2015)
  
 +==== Hardware Notes ====
 +  * The stage cannot go arbitrarily slow and still be perfectly smooth.  ASI's standard "scan optimized" stages use 16 TPI leadscrews that have a max speed of ~1.75 mm/s and a speed setting quantum of 172 nm/s.  To a fair approximation, the slowest speed that is relatively smooth is ~1000x slower than the max speed or about 12x the quantum.  So the "minimum smooth" speed is 1.8 um/sec for 16 TPI leadscrews.
 +
 +==== Post-Processing Notes ====
 +  *  As noted above the data needs to be deskewed before it can be used/viewed in a conventional way.  On-the-fly deskewing for visualizing test acquisitions and also for post-processing real acquisitions is in the plugin as of December 2016; the algorithm is fundamentally the same as the ImageJ macro but hopefully easier to use because the metadata provides most parameters (but background correction must be performed separately).
 +  * Unfortunately deskewing can blow up the size of (uncompressed) data files because a lot of empty space is added.
 +  * An alternative to deskewing is reslicing.  This can be as simple as reshuffling pixels in memory if the stage translation speed is chosen to be an integer number of pixels. Reslicing along the stage direction spreads the optical low-resolution between 2 axes but reduces the empty regions created by deskewing, particularly for long stage-scan stacks.  The following diagram hopefully helps:{{ :hardware:deskew_vs_reslice.jpg?600 |}}
 +  * It would be highly desirable for analysis and viewing software to be able to handle the deskew operation in memory and on the fly.  To date the only software known with this capability is FIJI's Big Data Viewer. Lobby your favorite software provider for this and make life better for everyone (many light sheet microscopes including lattice light sheet also require the same deskew step).
 +  * Micro-Manager 2.0 has a helpful ability to reslice data (an "Image Processor" added mid-2023) which is helpful especially for stage scanning data. The GPU version of the algorithm can operate on datasets up to 1/4 of the GPU memory, e.g. 2 GB datasets can be processed on a GPU with 8 GB of working memory.  The GPU version does interpolation and the CPU version is faster but doesn't do any interpolation.
 ==== Micro-Manager Implementation Notes ==== ==== Micro-Manager Implementation Notes ====
-  * The spacing between slices is specified in the plugin, so the stage moves sqrt(2) times that distance per slice. There may be very small rounding errors (which would come out in normal post-processing) but the metadata should contain the actual slice spacing instead of the user-specified one. +  * The spacing between slices is specified in the plugin, so the stage moves sqrt(2) times that distance per slice for 45 degree geometry (the actual angle can be set as of March 2018). There may be very small rounding errors (which would come out in normal post-processing) but the metadata should contain the actual slice spacing instead of the user-specified one.  Rounding errors were reduced as of the 2018-03-15 nightly build
-  * The initial X position should be the center of the dataset and the stage should return there afterwards.+  * The initial X position is used as the center X position of the dataset and the stage should return there afterwards
 +  * For large samples there is a YZ grid creator (as of March 2018) so multiple strips can be collected covering a sample.  The same GUI allows you to create a "grid" in X where instead of specifying slice spacing you specify start, stop, and number of slices.
   * There are three modes implemented, "stage scan", "stage scan interleaved", and "stage scan unidirectional"; see the [[docs:manual#spim_mode|manual's description]] of the different SPIM acquisition modes.   * There are three modes implemented, "stage scan", "stage scan interleaved", and "stage scan unidirectional"; see the [[docs:manual#spim_mode|manual's description]] of the different SPIM acquisition modes.
-  * As noted above the data needs to be deskewed before it can be used/viewed in a conventional way.  On-the-fly deskewing for visualizing test acquisitions and also for post-processing real acquisitions is in the plugin as of December 2016; the algorithm is fundamentally the same as the ImageJ macro and may not be quite as accurate as a full affine transformation.+