Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision Both sides next revision
hardware:objectives [2020/07/20 12:50]
Jon Daniels [Mechanical Angle] change 54-10-12 to 22deg mechanical
hardware:objectives [2020/07/28 12:03]
Jon Daniels [4f spacing]
Line 3: Line 3:
 The choice of light sheet objectives is limited because they must be co-focused without bumping into each other.  See more details below on this page in the section [[hardware:objectives#mechanical_angle|Mechanical angle]]. The choice of light sheet objectives is limited because they must be co-focused without bumping into each other.  See more details below on this page in the section [[hardware:objectives#mechanical_angle|Mechanical angle]].
  
-The most commonly-used objectives for (symmetric) diSPIM are 40x water-dipping objectives with a NA of 0.8 (Nikon CFI Apo 40XW NIR).  Other possibilities include the Olympus 20x/0.5 water (UMPLFLN20XW) ((By the mechanical drawings the Olympus 20x/0.5 water objectives will exactly touch when co-focused, but in practice it seems the working distance is slightly longer than specified so the spacing of the objectives at the tip is similar to the Nikon 40x/0.8 pair.)) and the Nikon 10x/0.3 water (CFI Plan Fluor 10XW). ASI and Special Optics have co-developed two different [[http://asiimaging.com/docs/cleared_tissue_objective|multi-immersion objectives]] designed originally for cleared tissue (but useful in any media) that is suitable for the diSPIM geometry.  The first with nominal NA 0.4 can image cleared tissue up to 5 mm deep in slab form or within a 12 mm spherical envelope, and the second with nominal NA 0.7 can go 2 mm deep or 10 mm spherical envelope.+The most commonly-used objectives for (symmetric) diSPIM are 40x water-dipping objectives with a NA of 0.8 (Nikon CFI Apo 40XW NIR).  Other possibilities include the Olympus 20x/0.5 water (UMPLFLN20XW) ((By the mechanical drawings the Olympus 20x/0.5 water objectives will exactly touch when co-focused, but in practice it seems the working distance is slightly longer than specified so the spacing of the objectives at the tip is similar to the Nikon 40x/0.8 pair.)) and the Nikon 10x/0.3 water (CFI Plan Fluor 10XW).
  
-Single-sided systems (iSPIM) have much more flexibility because the illumination objective can be a low-NA long-WD objective. A popular pair for high-resolution imaging is the same objective pair as used on the Lattice light sheet, specifically the Nikon 25x/1.1 objective paired with Special Optics 54-10-7 which is 28.6x/0.66.+ASI and Special Optics have co-developed two [[http://asiimaging.com/docs/cleared_tissue_objective|multi-immersion objectives]] designed originally for cleared tissue (but useful in any media) suitable for the diSPIM geometry.  The 54-10-12 with nominal NA 0.4 can image cleared tissue up to 5 mm deep in slab form or within a 12 mm spherical envelope, and the 54-12-8 with nominal NA 0.7 can go 2 mm deep or 10 mm spherical envelope. 
 + 
 +Single-sided systems (iSPIM) have much more flexibility because the illumination objective can be a low-NA long-WD objective. A popular pair for high-resolution imaging is the same objective pair as used on the lattice light sheet, specifically the Nikon 25x/1.1 objective paired with Special Optics 54-10-7 which is 28.6x/0.66.
  
 ==== Close-up Drawings ==== ==== Close-up Drawings ====
Line 35: Line 37:
 ==== Mechanical Angle ==== ==== Mechanical Angle ====
  
-For traditional light sheet microscopy with two orthogonal objective lenses, the objectives have to be able to co-focus before they mechanically bump ((For low-NA illumination you can sometimes extend the working distance a bit of the illumination objective by introducing diverging rays into its back aperture, but this is usually only a small win.)).  Regardless of working distance, the most important/fundamental factor in whether or not two objectives can be co-focused orthogonally is simply whether the sum of their mechanical half-angles is less than 90° ((However if the working distance of one is very long then perhaps they can co-focus with only use of the optical angle.)).  For any objective lens, the mechanical angle must be at least as big as the optical angle, i.e. it must be at least big enough to capture the cone of rays corresponding to its numerical aperture (NA).  The mechanical angle is computed as arctan(dia/2/WD) where dia is the diameter of the first surface (assuming the rest of the objective lens fall inside the line from the focal plane to this first surface as is usually the case).  The optical (half) angle is computed as arcsin(NA/RI) where RI is the medium refractive index.  Some objective lenses have mechanical angles only barely larger than the lower bound optical angle, but others are much less efficient in a mechanical/bulkiness sense.+For traditional light sheet microscopy with two orthogonal objective lenses, the objectives have to be able to co-focus before they mechanically bump ((For low-NA illumination you can sometimes extend the working distance a bit of the illumination objective by introducing diverging rays into its back aperture, but this is usually only a small win.)).  Regardless of working distance, the most important/fundamental factor in whether or not two objectives can be co-focused orthogonally is simply whether the sum of their mechanical half-angles is less than 90° ((However if the working distance of one is very long then perhaps they can co-focus with only use of the optical angle.)).  For any objective lens, the mechanical angle must be at least as big as the optical angle, i.e. it must be at least big enough to capture the cone of rays corresponding to its numerical aperture (NA) across the entire field of view.  The mechanical angle is computed as arctan(dia/2/WD) where dia is the diameter of the first surface (assuming the rest of the objective lens fall inside the line from the focal plane to this first surface as is usually the case).  The optical (half) angle is computed as arcsin(NA/RI) where RI is the medium refractive index.  Some objective lenses have mechanical angles only barely larger than the lower bound optical angle, but others are much less efficient in a mechanical/bulkiness sense.
  
 A detailed overview and helpful table of many (more) objective lenses can be found in Supplementary Note 6 in the Power/Huisken review paper ([[https://media.nature.com/original/nature-assets/nmeth/journal/v14/n4/extref/nmeth.4224-S1.pdf|link to supplemental]]). A detailed overview and helpful table of many (more) objective lenses can be found in Supplementary Note 6 in the Power/Huisken review paper ([[https://media.nature.com/original/nature-assets/nmeth/journal/v14/n4/extref/nmeth.4224-S1.pdf|link to supplemental]]).
Line 56: Line 58:
 | SO 54-10-12 17x/0.4 MI  |  16°                |  22°              | ASI multi-immersion #1 (RI 1.33 - 1.56, nominal 1.45)            | | SO 54-10-12 17x/0.4 MI  |  16°                |  22°              | ASI multi-immersion #1 (RI 1.33 - 1.56, nominal 1.45)            |
 | SO 54-12-8 24x/0.7 MI    29°                |  36°              | ASI multi-immersion #2 (RI 1.33 - 1.56, nominal 1.45)            | | SO 54-12-8 24x/0.7 MI    29°                |  36°              | ASI multi-immersion #2 (RI 1.33 - 1.56, nominal 1.45)            |
 +| SO 54-9-4 52x/1.15 MI    52°                |  57°              | (preliminary) ASI multi-immersion #3 (RI 1.33 - 1.56, nominal 1.45)            |
  
  
Line 64: Line 67:
 Different objectives have different positions of their back focal plane (BFP).  Olympus lists the BFP position of their objectives and so does ASI.  Nikon considers the BFP position confidential but it can be measured empirically. Different objectives have different positions of their back focal plane (BFP).  Olympus lists the BFP position of their objectives and so does ASI.  Nikon considers the BFP position confidential but it can be measured empirically.
  
-Most often ASI installes spacers between the scanner tube lens body and the Cube III containing the dichroic to adjust the total space, which is the easiest approach ifthe same objectives are always used.  ASI makes a 0-30mm continuously adjustable spacer which is useful if you are switching between objectives with different BFP positions or if you need to exactly tune the spacing (e.g. for using the virtual slit approach where the camera's rolling shutter is exactly synchronized with the motion of the beam).+Most often ASI installs spacers between the scanner tube lens body and the Cube III containing the dichroic to adjust the total space, which is the easiest approach if the same objectives are always used.  ASI makes a 0-30mm continuously adjustable spacer which is useful if you are switching between objectives with different BFP positions or if you need to exactly tune the spacing (e.g. for using the virtual slit approach where the camera's rolling shutter is exactly synchronized with the motion of the beam).
  
 The approximate spacers are listed here: The approximate spacers are listed here: