 ===== Comparison of diSPIM With Confocal ===== ===== Comparison of diSPIM With Confocal =====
  
-The following comparison was written with diSPIM in mind, but the main points apply to all types of SPIM.+The following comparison was written with diSPIM in mind, but the main points apply to all types of SPIM/LSFM.
  
-Spinning disk confocal (SDCM) is just a massively parallel confocal implementation so it's faster but otherwise has the some characteristics as laser scanning confocal.+Spinning disk confocal (SDCM) is just a massively parallel confocal implementation so it's faster but otherwise has the same characteristics as laser scanning confocal.
  
-  * **Light dose:** Confocal has much more photobleaching/phototoxicity than SPIM (i.e. SPIM uses the excitation light much more efficiently).  Low light dose is the main motivation for most SPIM users. +  * **Light dose:** diSPIM has much lower dose and hence less photobleaching/phototoxicity than confocalReducing light dose is the main motivation for most SPIM users. 
-      * In general for X slices you need X times as much light for confocal as you do for SPIM. +      * A rule of thumb is that you need X times as much light for confocal as you do for SPIM where X is the number of slices in your stack
-      * [[http://www.nature.com/nbt/journal/v31/n11/fig_tab/nbt.2713_SV2.html|Supplementary Video 2 of Wu et al.]] is a good comparison of imaging with diSPIM and SDCM; note how quickly the SDCM kills the cells being imaged+      * [[http://www.nature.com/nbt/journal/v31/n11/fig_tab/nbt.2713_F3.html|Figure 3]] and [[http://www.nature.com/nbt/journal/v31/n11/fig_tab/nbt.2713_SV2.html|Supplementary Video 2]] of Wu et al. offer bleaching comparisons of imaging with diSPIM and SDCM.  Other papers such as [[http://dx.doi.org/10.1002/jbio.201200144|this one]] offer comparisons of light sheet with confocal which should apply to diSPIM
-  * **Z resolution:** diSPIM after registration/fusion has best Z-resolution.  Next best is confocal, which in turn has better Z resolution than single-view SPIM. +  * **Z resolution:** diSPIM after registration/fusion has best Z-resolution (same as XY resolution).  Next best is confocal, which in turn has better Z resolution than single-view SPIM
-  * **XY resolution:** comparable between confocal and SPIM. +      * Using Bessel beams to create the sheet can improve Z-resolution for single-view SPIM ("optical sectioning"), but won't meaningfully change dual view resolution after registration/fusion
-  * **Speed:** SPIM (per view) and SDCM have comparable speed (assuming SDCM has higher laser intensity to make up for the ~3% open area of the spinning disks).+  * **XY resolution:** confocal and SPIM are comparable
 +  * **Speed:** SPIM (per view) and SDCM have comparable speed if SDCM laser intensity is increased to compensate for the ~3% open area of the disks
 +      * Camera readout speed bounds the maximum achievable frame rate.  For example, 512 pixels high ROI is 2.5 ms readout time for sCMOS.  Allowing 2.5 ms illumination time results in 5 ms total per image or 200 frames per second.  Using a cylindrical lens instead of a scanned light sheet can reduce camera overhead.