Line 171: Line 171:
  
 Currently several modes are supported: Currently several modes are supported:
-  * Synchronous piezo/slice scan: standard use, the piezo and sheet move together through the sample +  * Synchronous piezo/slice scan: standard piezo use, the piezo and sheet move together through the sample 
-  * No scan (fixed sheet): neither the piezo nor light sheet moves; useful for characterizing vibration in the system+  * No scan (fixed sheet): none of the piezo nor light sheet nor the stage moves (only the light sheet generation using the sheet axis); useful for characterizing vibration in the system
   * Stage scan: uses the XY stage to move the sample through a fixed light sheet; for two-sided acquisitions, one view is collected in the first pass, and the second view is collected as the stage moves “back” for a second pass; see the document on stage scanning for details and requirements   * Stage scan: uses the XY stage to move the sample through a fixed light sheet; for two-sided acquisitions, one view is collected in the first pass, and the second view is collected as the stage moves “back” for a second pass; see the document on stage scanning for details and requirements
   * Stage scan interleaved: same as stage scan except that the stage moves half as fast and the two light paths alternate, meaning both views are collected in the same pass; use caution if you are using a galvo to switch the laser between the sides, this can overstress some galvos   * Stage scan interleaved: same as stage scan except that the stage moves half as fast and the two light paths alternate, meaning both views are collected in the same pass; use caution if you are using a galvo to switch the laser between the sides, this can overstress some galvos
-  * Stage scan unidirectional: same as stage scan except the second view is taken while scanning in the same direction as the first view +  * Stage scan unidirectional: same as stage scan except the second view is taken while scanning in the same direction as the first view with a rewind step between.  For 1-sided acquisitions stage scan and stage scan unidirectional are equivalent.  For 2-sided acquisitions the stage scan unidirectional will take slightly longer (the rewind step is done at 75% of the max stage speed) but makes data processing easier. 
-  * Slice scan only: suppresses movement of piezo but moves the light sheet; useful for characterizing light sheet thickness +  * Slice scan only: suppresses movement of piezo but moves the light sheet; useful for characterizing light sheet thickness and for autofocus 
-  * Piezo scan only: the light sheet stays stationary but the piezo moves; useful for establishing axial PSF+  * Piezo scan only: the light sheet stays stationary but the piezo moves; useful for establishing axial PSF and for autofocus