| ===== Mount the SPIM Head ===== | ===== Mount the SPIM Head ===== | ||
| {{ youtube> | {{ youtube> | ||
| - | **Tools**: | + | **Tools**: |
| The video is cued up to show the process for mounting a diSPIM onto an ASI RAMM frame, but the principles apply to other scopes as well.\\ | The video is cued up to show the process for mounting a diSPIM onto an ASI RAMM frame, but the principles apply to other scopes as well.\\ | ||
| Line 18: | Line 18: | ||
| ===== Red Bracket ===== | ===== Red Bracket ===== | ||
| - | **Tools**: | + | **Tools**: |
| RAMM frames are shipped with a red bracket supporting the inverted scope; it must be removed. For detailed instructions with pictures, please visit [[http:// | RAMM frames are shipped with a red bracket supporting the inverted scope; it must be removed. For detailed instructions with pictures, please visit [[http:// | ||
| Line 27: | Line 27: | ||
| ===== Scanners ===== | ===== Scanners ===== | ||
| {{ youtube> | {{ youtube> | ||
| - | **Tools**: | + | **Tools**: |
| The video is cued up to show the process for mounting the scanners.\\ | The video is cued up to show the process for mounting the scanners.\\ | ||
| Line 36: | Line 36: | ||
| ===== Orientation ===== | ===== Orientation ===== | ||
| {{ youtube> | {{ youtube> | ||
| - | **Tools**: | + | **Tools**: |
| The video is cued up to show the process for mounting and orienting the cameras.\\ | The video is cued up to show the process for mounting and orienting the cameras.\\ | ||
| Line 64: | Line 64: | ||
| ===== Filters and Mirrors ===== | ===== Filters and Mirrors ===== | ||
| {{ youtube> | {{ youtube> | ||
| - | **Tools**: ASI ring tool, small Phillips head screwdriver, | + | **Tools**: ASI ring tool, small Phillips head screwdriver, |
| The video is cued up to show the process for installing an excitation filter, emission filter, and dichroic filter.\\ | The video is cued up to show the process for installing an excitation filter, emission filter, and dichroic filter.\\ | ||
| Line 70: | Line 70: | ||
| <WRAP left round info 50%> | <WRAP left round info 50%> | ||
| - | Notice: It is very easy to warp dichroic mirrors; only lightly tighten the retaining clip of the D-CUBE. | + | Notice: It is easy to warp dichroic mirrors |
| </ | </ | ||
| \\ \\ \\ \\ \\ | \\ \\ \\ \\ \\ | ||
| Line 106: | Line 106: | ||
| ===== Stage limits ===== | ===== Stage limits ===== | ||
| - | **Tools**: | + | **Tools**: |
| Setting the limit magnets on the motorized stages prevents crashes that can break coverslips or otherwise damage equipment; hardware limits are the most foolproof way to prevent crashes.((Limits may also be defined in software. Software-defined limits persist when the controller is powered down, but are lost when the firmware is updated or reset.)) The limit magnets move with the stage plates; when a magnet approaches a Hall effect sensor affixed in the stage body, the firmware detects a limit condition and stops the stage. The polarity of limit magnets relative to the stage is essential to their function. | Setting the limit magnets on the motorized stages prevents crashes that can break coverslips or otherwise damage equipment; hardware limits are the most foolproof way to prevent crashes.((Limits may also be defined in software. Software-defined limits persist when the controller is powered down, but are lost when the firmware is updated or reset.)) The limit magnets move with the stage plates; when a magnet approaches a Hall effect sensor affixed in the stage body, the firmware detects a limit condition and stops the stage. The polarity of limit magnets relative to the stage is essential to their function. | ||
| Line 116: | Line 116: | ||
| ===== Single axis ===== | ===== Single axis ===== | ||
| - | {{ youtube> | + | {{ youtube> |
| The video is cued up to demonstrate setting the F-axis limit magnet.\\ | The video is cued up to demonstrate setting the F-axis limit magnet.\\ | ||
| Line 157: | Line 157: | ||
| ===== XY-axes ===== | ===== XY-axes ===== | ||
| {{ youtube> | {{ youtube> | ||
| - | **Tools**: | + | **Tools**: |
| The video is cued up to demonstrate setting the XY-axes limit magnets; note that we now recommend setting the X-axis magnets first.\\ | The video is cued up to demonstrate setting the XY-axes limit magnets; note that we now recommend setting the X-axis magnets first.\\ | ||
| Line 186: | Line 186: | ||
| ===== Beam position ===== | ===== Beam position ===== | ||
| - | We are concerned with getting both the beam position and angle correct. Fortunately, | + | We are concerned with getting both the beam position and angle correct, along with getting both objectives " |
| < | < | ||
| ^ Adjustment | ^ Adjustment | ||
| - | | Tilt of dichroic mirror | + | | Tilt of dichroic mirror |
| - | | Tilt of camera mirror | + | | Tilt of camera mirror |
| - | | Position of upper Z stage | No | No | Angle varies if scanners are not connected to dichroic tubes | + | | Z position of SPIM head | No | No | |
| | Objective bushing (focus) | | Objective bushing (focus) | ||
| | Linear objective adjustment | | Linear objective adjustment | ||
| - | | < | + | | < |
| - | | < | + | | < |
| </ | </ | ||
| Line 248: | Line 248: | ||
| **Outline**\\ | **Outline**\\ | ||
| + | * Phase 0 (coarse alignment by eye as described above) | ||
| * Phase 1 (iterative) | * Phase 1 (iterative) | ||
| - | * 1. Co-align SPIM objectives to focus and center beams. | + | * 1.1 Co-align SPIM objectives to focus and center beams. |
| - | * 2. Tilt dichroic mirrors for horizontal | + | * 1.2 Tilt dichroic mirrors for __horizontal__ |
| - | * 3. Tilt imaging mirrors to center epi spots. | + | * 1.3 Tilt imaging mirrors to center epi spots. |
| + | * 1.4 Adjust lateral position of objective. | ||
| + | * 1.5 Go back to 1.1 and repeat until no adjustments needed | ||
| * Phase 2 (may have to be repeated for initial alignment, and again when scanners or cameras are changed or disturbed) | * Phase 2 (may have to be repeated for initial alignment, and again when scanners or cameras are changed or disturbed) | ||
| - | * 4. Focus collimators to center beam waists in FOV. | + | * 2.1 Focus collimators to center beam waists in FOV. |
| - | * 5. Rotate scanners for __uniform__ focus of the sheets. | + | * 2.2 Rotate scanners for __uniform__ focus across |
| - | * 6. Rotate cameras for vertical sheets from epi view. | + | * 2.3 Rotate cameras for vertical sheets from epi view. |
| - | * Phase 3 (performed just once for initial alignment and again for objective | + | * Go back to 1.1 |
| - | * 7. Establish the coverslip location. | + | * Phase 3 (performed just once for initial alignment and again for objective |
| - | * 8. Align the inverted scope with the SPIM objectives. | + | * 3.1 Establish the coverslip location. |
| - | * 9. Cross-calibrate the piezo and light sheet movement. | + | * 3.2 Align the inverted scope with the SPIM objectives. |
| + | * 3.3 Cross-calibrate the piezo and light sheet movement. | ||
| + | * Phase 4 (performed for any new sample) | ||
| + | * 4.1 Check that light sheet and detection plane are coincident; adjust offset in plugin if needed. | ||
| ==== Phase 1 ==== | ==== Phase 1 ==== | ||
| - | Repeat the steps of Phase 1 steps multiple times (after the first iteration, in no particular order) | + | Repeat the steps of Phase 1 steps multiple times as you spiral in on alignment. |
| === Cofocus SPIM objectives === | === Cofocus SPIM objectives === | ||
| {{ youtube> | {{ youtube> | ||
| - | **Adjust**: OBLPA (<imgref piezo_with_bushing> | + | **Configure** the [[docs: |
| - | **End goal**: Both beams in focus with the epi-spot sitting on the beam waist (roughly at first, as in <imgref BeginAlign>, | + | select |
| - | The video is cued up to demonstrate cofocusing | + | use the "On tab activate" |
| - | In the [[docs:mm_dispim_plugin_user_guide# | + | **Adjust**: |
| + | * both objective bushings, and | ||
| + | * the knob on the OBLPA (<imgref piezo_with_bushing> | ||
| + | **The goal** is to have both beams in focus (roughly at first, as in <imgref BeginAlign>, | ||
| + | The video is cued up to demonstrate cofocusing the SPIM objectives using Micro-Manager.\\ | ||
| This is best done by alternating between steering and focus adjustments, | This is best done by alternating between steering and focus adjustments, | ||
| Line 347: | Line 357: | ||
| **Adjust**: Rotation of the camera (and camera' | **Adjust**: Rotation of the camera (and camera' | ||
| **End goal**: Epi-view of the sheets produces a vertical line in the camera image\\ | **End goal**: Epi-view of the sheets produces a vertical line in the camera image\\ | ||
| - | **Tools**: | + | **Tools**: |
| The video is cued up to demonstrate adjustment of the camera' | The video is cued up to demonstrate adjustment of the camera' | ||
| - | If the epi-view of the sheet is not vertical, | + | If the epi-view of the sheet is not vertical, |
| If you make any changes in Phase 2, return to the start of Phase 1. | If you make any changes in Phase 2, return to the start of Phase 1. | ||