| **Outline**\\  | **Outline**\\  | ||
| + | * Phase 0 (coarse alignment by eye as described above) | ||
| * Phase 1 (iterative) | * Phase 1 (iterative) | ||
| - | * 1. Co-align SPIM objectives to focus and center beams. | + | * 1.1 Co-align SPIM objectives to focus and center beams. | 
| - |     * 2. Tilt dichroic mirrors for horizontal  | + |     * 1.2 Tilt dichroic mirrors for __horizontal__  | 
| - | * 3. Tilt imaging mirrors to center epi spots. | + | * 1.3 Tilt imaging mirrors to center epi spots. | 
| + | * 1.4 Adjust lateral position of objective. | ||
| + | * 1.5 Go back to 1.1 and repeat until no adjustments needed | ||
| * Phase 2 (may have to be repeated for initial alignment, and again when scanners or cameras are changed or disturbed) | * Phase 2 (may have to be repeated for initial alignment, and again when scanners or cameras are changed or disturbed) | ||
| - | * 4. Focus collimators to center beam waists in FOV. | + | * 2.1 Focus collimators to center beam waists in FOV. | 
| - | * 5. Rotate scanners for __uniform__ focus of the sheets. | + |     * 2.2 Rotate scanners for __uniform__ focus across  | 
| - | * 6. Rotate cameras for vertical sheets from epi view. | + | * 2.3 Rotate cameras for vertical sheets from epi view. | 
| - |   * Phase 3 (performed just once for initial alignment and again for objective  | + | * Go back to 1.1 | 
| - | * 7. Establish the coverslip location. | + |   * Phase 3 (performed just once for initial alignment and again for objective  | 
| - | * 8. Align the inverted scope with the SPIM objectives. | + | * 3.1 Establish the coverslip location. | 
| - | * 9. Cross-calibrate the piezo and light sheet movement. | + | * 3.2 Align the inverted scope with the SPIM objectives. | 
| + | * 3.3 Cross-calibrate the piezo and light sheet movement. | ||
| + | * Phase 4 (performed for any new sample) | ||
| + | * 4.1 Check that light sheet and detection plane are coincident; adjust offset in plugin if needed. | ||