Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision
Previous revision
docs:modular_manual [2020/08/18 14:54]
Jon Daniels [Beam position]
docs:modular_manual [2022/10/25 22:47] (current)
Jon Daniels [Fine alignment]
Line 70: Line 70:
  
 <WRAP left round info 50%> <WRAP left round info 50%>
-Notice: It is very easy to warp dichroic mirrors; only lightly tighten the retaining clip of the D-CUBE.  +Notice: It is easy to warp dichroic mirrors enough to induce aberrations; only lightly tighten the retaining clip of the D-CUBE.  
 </WRAP> </WRAP>
 \\ \\ \\ \\ \\  \\ \\ \\ \\ \\ 
Line 248: Line 248:
  
 **Outline**\\  **Outline**\\ 
 +  * Phase 0 (coarse alignment by eye as described above)
   * Phase 1 (iterative)   * Phase 1 (iterative)
-    * 1. Co-align SPIM objectives to focus and center beams. +    * 1.Co-align SPIM objectives to focus and center beams. 
-    * 2. Tilt dichroic mirrors for horizontal beams __uniformly__ in/out of focus.  +    * 1.Tilt dichroic mirrors for __horizontal__ beams __uniformly__ in/out of focus.  
-    * 3. Tilt imaging mirrors to center epi spots.+    * 1.Tilt imaging mirrors to center epi spots. 
 +    * 1.4 Adjust lateral position of objective. 
 +    * 1.5 Go back to 1.1 and repeat until no adjustments needed
   * Phase 2 (may have to be repeated for initial alignment, and again when scanners or cameras are changed or disturbed)   * Phase 2 (may have to be repeated for initial alignment, and again when scanners or cameras are changed or disturbed)
-    * 4. Focus collimators to center beam waists in FOV. +    * 2.Focus collimators to center beam waists in FOV. 
-    * 5. Rotate scanners for __uniform__ focus of the sheets. +    * 2.Rotate scanners for __uniform__ focus across the sheets. 
-    * 6. Rotate cameras for vertical sheets from epi view. +    * 2.Rotate cameras for vertical sheets from epi view. 
-  * Phase 3 (performed just once for initial alignment and again for objective or sample changes+    * Go back to 1.1 
-    * 7. Establish the coverslip location. +  * Phase 3 (performed just once for initial alignment and again for objective change
-    * 8. Align the inverted scope with the SPIM objectives. +    * 3.Establish the coverslip location. 
-    * 9. Cross-calibrate the piezo and light sheet movement.+    * 3.Align the inverted scope with the SPIM objectives. 
 +    * 3.Cross-calibrate the piezo and light sheet movement
 +  * Phase 4 (performed for any new sample)  
 +    * 4.1 Check that light sheet and detection plane are coincident; adjust offset in plugin if needed.
  
 ==== Phase 1 ==== ==== Phase 1 ====