Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision
Previous revision
hardware:objectives [2024/03/07 00:37]
Jon Daniels [Mechanical Angle]
hardware:objectives [2024/11/01 22:02] (current)
Jon Daniels [Mechanical Angle]
Line 7: Line 7:
 ASI and Special Optics have co-developed two [[http://asiimaging.com/docs/cleared_tissue_objective|multi-immersion objectives]] designed originally for cleared tissue (but useful in any media) suitable for the diSPIM geometry.  The 54-10-12 with nominal NA 0.4 can image cleared tissue up to 5 mm deep in slab form or within a 12 mm spherical envelope, and the 54-12-8 with nominal NA 0.7 can go 2 mm deep or 10 mm spherical envelope. ASI and Special Optics have co-developed two [[http://asiimaging.com/docs/cleared_tissue_objective|multi-immersion objectives]] designed originally for cleared tissue (but useful in any media) suitable for the diSPIM geometry.  The 54-10-12 with nominal NA 0.4 can image cleared tissue up to 5 mm deep in slab form or within a 12 mm spherical envelope, and the 54-12-8 with nominal NA 0.7 can go 2 mm deep or 10 mm spherical envelope.
  
-Single-sided systems (iSPIM) have much more flexibility because the illumination objective can be a low-NA long-WD objective. A popular pair for high-resolution imaging is the same objective pair as used on the lattice light sheet, specifically the Nikon 25x/1.1 objective paired with Special Optics 54-10-7 which is 28.6x/0.66.+Single-sided systems (iSPIM) have much more flexibility because the illumination objective can be a low-NA long-WD objective. A popular pair for high-resolution imaging is the same objective pair as used on the lattice light sheet, specifically the Nikon 25x/1.1 objective paired with either the Special Optics 54-10-7 which is 28.6x/0.66, or else paired with the Thorlabs 20x/0.6.
  
 ==== Close-up Drawings ==== ==== Close-up Drawings ====
Line 51: Line 51:
 | Olympus 20x/0.5 W        22°                |  45°              | fits better than expected (WD is probably a bit more than spec)  | | Olympus 20x/0.5 W        22°                |  45°              | fits better than expected (WD is probably a bit more than spec)  |
 | Olympus 20x/1.0 W        49°                |  52°              |   | | Olympus 20x/1.0 W        49°                |  52°              |   |
 +| Nikon 20x/1.0 W          49°                |  53°              |                                           |
 +| Leica 20x/1.0 W          49°                |  52°              |                                           |
 +| Zeiss 20x/1.0 W          49°                |  51°              | 421452-9800-00                            |
 | Nikon 25x/1.1 W          56°                |  58°              | usual lattice detection                                          | | Nikon 25x/1.1 W          56°                |  58°              | usual lattice detection                                          |
 | Olympus 60x/1.0 W        49°                |  53°              | sometimes oSPIM detection           | | Olympus 60x/1.0 W        49°                |  53°              | sometimes oSPIM detection           |
Line 56: Line 59:
 | Nikon 60x/1.0 W          49°                |  57°              |            | | Nikon 60x/1.0 W          49°                |  57°              |            |
 | SO 54-10-7 29x/0.66 W    30°                |  30°              | traditional lattice illumination                                 | | SO 54-10-7 29x/0.66 W    30°                |  30°              | traditional lattice illumination                                 |
-| TL20X-MPL 20x/0.6 W      27°                |  ~30°             | new (2020) cost-effective lattice illumination, only 7mm FN?!    |+| TL20X-MPL 20x/0.6 W      27°                |  ~30°             | new (2020) cost-effective lattice illumination, note only 7mm FN at full NA    |
 | Nikon 20x/1.0 glyc      |  43°                |  ~54°             | cleared tissue confocal used in light sheet (RI 1.44 - 1.50)     | | Nikon 20x/1.0 glyc      |  43°                |  ~54°             | cleared tissue confocal used in light sheet (RI 1.44 - 1.50)     |
 | Olympus 25x/1.0 glyc    |  43°                |  56°              | cleared tissue confocal used in light sheet (RI 1.41 - 1.52)     | | Olympus 25x/1.0 glyc    |  43°                |  56°              | cleared tissue confocal used in light sheet (RI 1.41 - 1.52)     |
-| SO 54-10-12 17x/0.4 MI  |  16°                |  22°              | ASI multi-immersion #1 (RI 1.33 - 1.56, nominal 1.45)            +| SO 54-10-12 17x/0.4 MI  |  16°                |  32°              | ASI multi-immersion #1 (RI 1.33 - 1.56, nominal 1.45), mechanical limitation in body   
-| SO 54-12-8 24x/0.7 MI    29°                |  36°              | ASI multi-immersion #2 (RI 1.33 - 1.56, nominal 1.45)            |+| SO 54-12-8 24x/0.7 MI    29°                |  36°              | ASI multi-immersion #2 (RI 1.33 - 1.56, nominal 1.45)   | 
 +| SO 57-12-19 10x/0.3 MI  |  11°                |  28°              | ASI multi-immersion #3 (RI 1.33 - 1.56, nominal 1.45), mechanical limitation in body  |
 | SO 54-9-4 52x/1.15 MI    52°                |  57°              | (preliminary) ASI multi-immersion #3 (RI 1.33 - 1.56, nominal 1.45)            | | SO 54-9-4 52x/1.15 MI    52°                |  57°              | (preliminary) ASI multi-immersion #3 (RI 1.33 - 1.56, nominal 1.45)            |
 +
 +
 +==== Working Distance and Imaging Depth ====
 +
 +The imaging depth of a pair of co-focused objectives into a flat pancake-like sample without hitting the top of the sample is proportional to the residual mechanical angle that the two objectives leave and also proportional to their working distance, with minor factor related to the relative mechanical angle/overall rotation.  There may also be second-order mechanical effects and such combinations should be modeled.  ASI has drawings of many different objective pairs and can prepare others.  Please inquire.