Differences
This shows you the differences between two versions of the page.
Both sides previous revision Previous revision Next revision | Previous revision | ||
hardware:objectives [2024/05/01 09:27] Jon Daniels [Mechanical Angle] |
hardware:objectives [2024/11/01 22:02] (current) Jon Daniels [Mechanical Angle] |
||
---|---|---|---|
Line 51: | Line 51: | ||
| Olympus 20x/0.5 W | | Olympus 20x/0.5 W | ||
| Olympus 20x/1.0 W | | Olympus 20x/1.0 W | ||
+ | | Nikon 20x/1.0 W | ||
+ | | Leica 20x/1.0 W | ||
+ | | Zeiss 20x/1.0 W | ||
| Nikon 25x/1.1 W | | Nikon 25x/1.1 W | ||
| Olympus 60x/1.0 W | | Olympus 60x/1.0 W | ||
Line 59: | Line 62: | ||
| Nikon 20x/1.0 glyc | 43° | ~54° | cleared tissue confocal used in light sheet (RI 1.44 - 1.50) | | | Nikon 20x/1.0 glyc | 43° | ~54° | cleared tissue confocal used in light sheet (RI 1.44 - 1.50) | | ||
| Olympus 25x/1.0 glyc | 43° | 56° | cleared tissue confocal used in light sheet (RI 1.41 - 1.52) | | | Olympus 25x/1.0 glyc | 43° | 56° | cleared tissue confocal used in light sheet (RI 1.41 - 1.52) | | ||
- | | SO 54-10-12 17x/0.4 MI | 16° | | + | | SO 54-10-12 17x/0.4 MI | 16° | |
- | | SO 54-12-8 24x/0.7 MI | + | | SO 54-12-8 24x/0.7 MI |
+ | | SO 57-12-19 10x/0.3 MI | 11° | 28° | ASI multi-immersion #3 (RI 1.33 - 1.56, nominal 1.45), mechanical limitation in body | ||
| SO 54-9-4 52x/1.15 MI | | SO 54-9-4 52x/1.15 MI | ||
+ | |||
+ | |||
+ | ==== Working Distance and Imaging Depth ==== | ||
+ | |||
+ | The imaging depth of a pair of co-focused objectives into a flat pancake-like sample without hitting the top of the sample is proportional to the residual mechanical angle that the two objectives leave and also proportional to their working distance, with minor factor related to the relative mechanical angle/ | ||