The Idea

The diSPIM is a flexible and easy-to-use implementation of Selective Plane Illumination Microscopy (SPIM) that allows for dual views (d) of the sample while mounted on an inverted (i) microscope. The diSPIM was co-developed by the lab of Hari Shroff at NIH/NIBIB and Applied Scientific Instrumentation (ASI). SPIM is also referred to as light sheet fluorescence microscopy or LSFM because it uses a sheet or plane of light to illuminate the sample perpendicular to the imaging direction.

The Implementation

The diSPIM has two (usually symmetric) optical paths for light sheet imaging. Two objectives are placed at right angles above a sample mounted horizontally. A light sheet is created from one objective and then imaged using the other objective. A stack of images is collected by moving the light sheet through the sample. For a few users, the 3D information from a single view or stack is sufficient (iSPIM). For dual-view systems, the role of the two objectives is reversed to collect another stack from a perpendicular direction, and then the two datasets can be computationally merged to yield a 3D dataset with isotropic resolution (the usual problem of poor axial resolution is overcome by information from the other view).

The diSPIM “head” can be mounted on various inverted microscopes including ASI's RAMM frame (shown here). diSPIM systems can be obtained from various system integrators. Various open-source and proprietary software packages are available for both data acquisition and data visualization. Most of the underlying microscope hardware is identical regardless of the system integrator and software used.

The choice of diSPIM objectives is limited because they must be co-focused without bumping into each other. The most commonly-used objectives for diSPIM are 40x water-dipping objectives with a NA of 0.8 (Nikon CFI Apo 40XW NIR).

Most often sCMOS cameras are used for SPIM imaging. There are working diSPIM systems with Hamamatsu Flash4, Andor Zyla, and PCO Edge cameras.

ASI makes a compact fiber-coupled 2D galvo or “scanner” which is an integral part of the system. It creates the light sheet by fast scanning in one axis and moves the sheet through the sample using the other axis.1) The output of the excitation laser (or laser launch) simply is fed into the scanner; it is helpful to have a 2×1 optical switch or dual-output laser launch so the excitation can all be steered to the scanner in the active light path.

For applications where environmental control is important, the diSPIM can easily be fitted with an incubator enclosure and appropriate equipment to keep samples alive and happy.

The bottom objective (the inverted microscope) most often has a lower-magnification objective and less expensive camera for locating the sample. Epi-illumination can easily be added.

The Benefits

Like other light sheet techniques, diSPIM illuminates only the focus plane and is thus ideal for imaging living cells and organisms because it minimizes photobleaching and phototoxicity effects. Compared with confocal or spinning disk systems, there is ~2x improvement in axial resolution, >10x reduction in photobleaching, and speed comparable to spinning disk. See a more detailed comparison with confocal.

A major advantage compared with most other light sheet implementations is that sample mounting is extremely simple, similar to an inverted microscope. Most commonly, specimens are placed on a 24x50mm coverslip which is held in a special chamber that holds the dipping media. See a more detailed comparison of light sheet methods.

Besides being handy to locate samples, the bottom objective can be used for photoactivation (including optogenetics) or other experimental techniques. It also can be used to provide a third independent view of the sample.

The diSPIM is a modular microscope so many variations and add-ons are possible depending on your specific needs. Various new features and improvements are are being explored by NIH researchers, Applied Scientific Instrumentation, and others. See a partial list of variations.

diSPIM systems are commercially available from multiple system integrators. Compared with other commercial light-sheet solutions they are inexpensive and flexible. Compared with custom-built SPIM/LSFM systems they are easy to obtain, use, and maintain. See also a more complete technical comparison of SPIM techniques.

This Wiki

This wiki is a place for diSPIM users (and future diSPIM users) to exchange information in a decentralized way. Please join the community and contribute your expertise!

Content is readable by anybody but to prevent abuse only registered users can edit. Register for an account here.

There is also a mailing list with searchable archives that can be accessed at lists.dispim.org. This mailing list hasn't been used as of June 2015.

1)
i.e. the diSPIM is a digital light sheet microscope or DLSM